Potentiometric sensor fabrication having 2D sarcosine memories and analytical features
Göster/ Aç
Erişim
info:eu-repo/semantics/closedAccessTarih
2016Yazar
Özkütük, Ebru BirlikEmir Diltemiz, Sibel
Avcı, Seyma
Uğurag, Deniz
Aykanat, Rabia Berna
Ersöz, Arzu
Say, Rıdvan
Üst veri
Tüm öğe kaydını gösterÖzet
In this study, a simple, rapid and sensitive method based on novel molecular imprinted polymeric sensor has been developed and validated for the determination of prostate cancer metabolite biomarker. The molecularly imprinted polymer (MIP) has been synthesized by emulsion polymerization, using sarcosine as template molecule, methacryloylamido histidine (MAH) as functional monomer and ethylene glycol dimethacrylate (EDMA) as cross-linking agent. The performance of the developed sarcosine sensor has been evaluated, and the results have indicated that a sensitive potentiometric sensor has been fabricated. The sarcosine sensor has showed high-selectivity, shorter response time (<2 min), wider linear range (10(-2)-10(-6) mM), lower detection limit (1.35 x 10(-7) mM), and satisfactory long-term stability (>5.5 months)
Kaynak
Materials Science & Engineering C-Materials For Biological ApplicationsCilt
69Koleksiyonlar
- Makale Koleksiyonu [1058]
- PubMed İndeksli Yayınlar Koleksiyonu [1498]
- Scopus İndeksli Yayınlar Koleksiyonu [8325]
- WoS İndeksli Yayınlar Koleksiyonu [7605]